Asian Journal of Chemistry; Vol. 25, No. 9 (2013), 4788-4790

http://dx.doi.org/10.14233/ajchem.2013.14105

Validated RP-HPLC Method for the Determination of Mycophenolate Mofetil in Tablet Dosage Forms T. VIJAYA BHASKARA REDDY, G. RAMU, M. SRAVAN KUMAR and C. RAMBABU* Department of Chemistry, Acharya Nagarjuna University, Nagarjuananagar-522 510, India *Corresponding author: E-mail: [email protected] (Received: 24 May 2012;

Accepted: 6 March 2013)

AJC-13072

A precise, accurate and more sensitive RP-HPLC method is developed to estimate the amount of mycophenolate mofetil in pure and tablets using Waters-Alliance HPLC system equipped with auto sampler and ultra-violet detector. The method is carried out on a Symmetry C18 (4.6 mm ID × 150 mm, 5 µm, Make: XTerra) with a mobile phase consisting of acetonitrile and potassium dihydrogen phosphate buffer of pH = 4.0 in the ratio 65:35 volume/volume at a flow rate of 0.7 mL/min. The detection of eluted components is carried out at a wavelength of 216 nm. The retention time of mycophenolate mofetil is found to be 2.647 min. The developed method is validated in terms of accuracy, precision, linearity, limit of detection, limit of quantization. The linearity limits, LOD and LOQ of the developed method are found to be 10-50, 0.052 and 0.171 µg mL-1, respectively. The developed method is found to be simple, fast and economic and hence it can be used as an alternative method in quality control. Key Words: RP-HPLC, Mycophenolate mofetil, Retention time, Linearity, Quality control.

INTRODUCTION Mycophenolate mofetil (MMF) is an immunosuppressant and prodrug of mycophenolic acid, used extensively in transplant medicine. It is a reversible inhibitor of inosine monophosphate dehydrogenase1 (IMPDH) in purine biosynthesis, more specifically guanine synthesis, which is necessary for the growth of T cells and B cells. Mycophenolate mofetil is also used in the treatment of autoimmune diseases, such as Behcet's disease, pemphigus vulgaris and systemic lupus erythematosus. The chemical name for mycophenolate mofetil is 2-morpholinoethyl (E)-6-(1,3-dihydro-4-hydroxy-6methoxy-7-methyl-3-oxo-5-isobenzofuranyl)-4-methyl-4hexenoate. Its empirical formula is C23H31NO7 with molecular weight 433.50 and soluble in acidic medium. It is available in the brand names such as CellCept, Myfortic. One UV-spectrophotometric method is reported 2 to estimate mycophenolate mofetil in bulk and pharmaceutical formulations at 250 nm in 0.1 N hydrochloric acid (pH 1.2) using acetate buffer (pH 4.5). A high-performance liquidchromatography method3 is developed and validated for the estimation of mycophenolate mofetil in tablets, capsules and nanoparticles by using photo diode array detector. A simultaneous determination of mycophenolic acid and valproic acid in human plasma by HPLC4 is reported based on derivatization by high-performance liquid chromatography with fluorescence

detection. A few LC-MS/MS5-10 methods have been reported for the determination of mycophenolate mofetil present in biological fluids in humans. A few HPLC11-14 are also reported in the literature to determine the amount of the drug mostly in biological matrixes. Most of the reported methods are concentrated on the biological fluids. From the extensive surrey of the literature it is found that a very few methods are present to determine the drug in formulations. Therefore the author has attempted to develop a more sensitive HPLC method for the assay of the chosen drug in pharmaceutical formulations. The LOD and LOQ values of the reported method are very much lower than that of that of the reported method3. More over the linearity limits for the proposed method are 10-50 µg mL-1 which is lower than the existing method of 120-800 µg mL-1 indicating that the method is more sensitive. The UV absorption spectrum (Fig. 1) of the drug gives three peaks at 216, 250 and 304 nm. It is found that the absorbance at 216 nm is nearly five times more than the one at 250 nm. Hence the all spectral measurements are made at 216 nm to achieve better sensitivity.

EXPERIMENTAL The separation is carried out on Waters-Alliance HPLC system equipped with auto sampler, binary gradient pump, dual wavelength UV-visible detector and symmetry C18 (4.6

Vol. 25, No. 9 (2013)

Validated RP-HPLC Method for the Determination of Mycophenolate Mofetil 4789

1.500

0.50

Abs

0.40

1.123

0.30 0.20

Abs

0.10 0.00

0.745

0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00 5.50 6.00 Minutes

3

Fig. 2. A typical RP-HPLC chromatogram of mycophenolate standard 30 µg mL-1 0.367 2

-0.010 200

1

Conc. 250

300

350

400

Wavelength (nm)

Fig. 1. UV Absorption spectrum of the mycophenolate mofetil

mm ID × 150 mm, 5 µm, Make: XTerra) column. Empower chromatography data software is used to get the reports. Analytical grade potassium dihydrogen phosphate and ortho phosphoric acid are purchased from Qualigens Fine Chemicals Ltd., Mumbai. Acetonitrile of HPLC grade is obtained from E. Merck. (India) Ltd., Mumbai. Mycophenolate mofetil of highest purity (99.8 %) is gifted by Dr. Reddy's Laboratories Ltd., Hyderabad. The commercially available mycophenolate mofetil tablets CellCept and Myfortic are purchased from the local market. Preparation of mobile phase: The mobile phase is prepared by adding of 650 mL acetonitrile (HPLC grade) to 350 mL of 0.7 % potassium dihydrogen phosphate buffer of pH 4.0 (the pH of the solution is adjusted to 4.0 by adding 1 % ortho phosphoric acid solution), sonicated and filtered through 0.45 µ membrane filter before use. Preparation of the mycophenolate standard and sample solution: Stock solution of the mycophenolate mofetil is prepared by dissolving accurately weighed 10 mg of mycophenolate standard (for preparation of test solution, weight of the sample powder equivalent to 10 mg of the mycophenolate is taken) in 7 mL of mobile phase in a 10 mL volumetric flask, sonicated, made up to the mark with the mobile phase and filtered through 0.45 µm membrane filter. A series dilute solutions ranging from 10-50 µg mL-1 are prepared from the stock solution in mobile phase. Chromatographic conditions: The mobile phase is pumped from the solvent reservoir into the column at a flow rate of 0.7 mL/min. The column is allowed to equilibrate for 0.5 h prior to the injection of 20 µL the standard. The detection of the components that are present in the elutent is monitored at 216 nm. The working standard solution of concentration 30 µg mL-1 is injected into 20 µL loop and the chromatogram is recorded and represented in Fig. 2. The retention time of the drug is found to be 2.647 min. The system suitable parameters such as tailing factor and number of theoretical plates are found to be within the limits and are reported in Table-1. Precision: The intra-day precision (concordance of the experimentally determined values) of a method is expressed in terms of statistical parameters such as standard deviation and RSD %. The RSD % is calculated from the standard deviation

30 µg mL-1

TABLE-1 SYSTEM SUITABILITY PARAMETERS Parameters Obtained value Acceptance value Tailing factor 1.6 Not more than 2.0 Theoretical plate Not less than 2639 count 2000

and mean area of five replicate measurements of the working standard solution of concentration 30 µg mL-1. Inter-day precision of the method is determined by carrying out the experiment on different days using same instrument and same column under similar chromatographic conditions. The RSD % value for the five replicate measurements is found to be 0.29 (Table-2). TABLE-2 PRECISION OF THE METHOD Intra-day precision Inter day precision Injection Area Injection Area Injection-1 4991365 Injection-1 4860869 Injection-2 4963809 Injection-2 4828763 Injection-3 5000570 Injection-3 4843661 Injection-4 4817340 Injection-4 4854142 Injection-5 4820237 Injection-5 4862617 Average 4918664 Average 4850011 Standard deviation 92177.1 Standard deviation 14015.1 RSD (%) 1.87 RSD (%) 0.29

Accuracy: The accuracy (closeness of the experimentally determined value to the most probable value or true value) of the method is determined by recovery experiments. The recovery studies are carried out at three different concentrations i.e., 50, 100 and 150 % of target concentration and presented in Table-3. TABLE-3 ACCURACY OF THE PROPOSED METHOD Conc Amount Amount Recovery Mean Area (%) added found (%) recovery 50 2908918 5.23 mg 5.31 mg 101.5 99.9% 100 5310069 9.8 mg 9.69 mg 99.9 150 7387647 13.6 mg 13.4 mg 99.1 Acceptance criteria: The per cent of recovery for each level should be between 98.0-102.0 %.

Linearity: A series of standard solutions (10-50 µg mL-1) are injected into the 20 µL loop and the chromatogram is recorded for each concentration. The calibration curve is constructed by plotting concentration against peak area and is represented in Fig. 3. The correlation between concentration of the drug and peak area is expressed in terms of correlation

4790 Reddy et al.

Asian J. Chem.

9000000

RESULTS AND DISCUSSION

8000000

y = 16351x + 78025 R2 = 0.9998

7000000 Peak area

6000000 5000000 4000000 3000000 2000000 1000000 0

0

10

20

30

40

50

60

Concentration of the drug (µ µg/mL)

Fig. 3. Linearity of the drug concentration against the area of the peak

coefficient. The slope and intercept are evaluated by linear regression analysis using linear least square method. The slope, intercept and correlation coefficient values are given in Table-4. TABLE-4 LINEARITY OF THE DRUG CONCENTRATION WITH PEAK AREA Concentration (µg/mL-1) Area of the peak 10 1723966 20 3355127 30 5093790 40 6723617 50 8098201 Slope 16351 Intercept 78025 Correlation coefficient 0.9998 LOD 0.052 µg/mL-1 LOQ 0.173 µg/mL-1 Acceptance criteria: Correlation coefficient should be not less than 0.999.

Limit of detection (LOD) and limit of quantification (LOQ): The limit of detection (LOD) is the smallest concentration of the analyte that gives the measurable response. The limit of quantification (LOQ) is the smallest concentration of the analyte, which gives response that can be accurately quantified. The experimentally determined values of LOD and LOQ are given in Table-4. LOD and LOQ are calculated using (3.3 × standard deviation)/slope of calibration curve and (10.0 × standard deviation)/slope of calibration curve, respectively. The LOD and LOQ of the solution are found to be 0.0521 and 0.171 µg mL-1. Robustness: Robustness of the proposed method is checked by making slight deliberate change in the experimental procedures. In the present method a deliberate change in the flow rate and mobile phase composition is made to evaluate the impact on the method. (Table-5).

UV absorption spectrum of mycophenolate is scanned from 200-400 nm with 2 nm band width which indicates three peaks at 216, 250 and 304 nm (Fig. 1). It is found that the absorbance at 216 nm is nearly five times more and hence we have selected 216 nm as the suitable wavelength for recording chromatograms in the proposed RP-HPLC method. The mixture of potassium dihydrogen phosphate solution and acetonitrile (35:65) is selected as mobile phase with a flow rate 0.7 mL/ min. The retention time is found to be 2.647 min. The drug obeys linearity in the range of 10-50 µg mL-1 and the correlation coefficient is found to be 0.9999. The developed method is found to be accurate and precise as indicated by recovery studies and RSD % not more than 2.0. Moreover LOD and LOQ are found to be less than the reported values. Recovery studies are performed at 50, 100 and 150 % concentration levels and the results are found to be satisfactory. The results indicate that the proposed method is found to be more sensitive than the existing methods. Conclusion The proposed RP-HPLC method for the estimation of mycophenolate mofetil in dosage forms is found to be more precise and sensitive. The per cent recovery of the drug (accuracy) indicates that the proposed method is selective. The drug is eluted in a reasonable retention time 2.647 min which indicates speed of the analysis, hence small portions of the solvents are sufficient for the complete analysis therefore the cost of the method is reduced.

ACKNOWLEDGEMENTS The authors are thankful to Pharma Train, Hyderabad for providing the laboratory facilities.

REFERENCES 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11.

TABLE-5 STUDY OF ROBUSTNESS OF THE PROPOSED METHOD Mobile Flow rate Plate Tailing Plate Tailing phase (mL/min) count factor count factor composition 0.6 mL/min 2444.1 1.7 10 % less 2557.3 1.2 0.7 mL/min 2639.8 1.6 Actual 2639.8 1.6 0.8 mL/min 2648.9 1.5 10 % high 2503.7 1.6 *Results for actual flow (0.7 mL/min) have been considered from assay standard. *Results for actual mobile phase composition (65:35 acetonitrile:buffer) have been considered.

12. 13. 14.

B. Fulton and A. Markham, Drugs, 51, 278 (1996). S. Verma, H. Gupta, O. Alam, P. Mullick, N. Siddiqui and S. Khan, J. Appl. Spectrosc., 76, 876 (2009). Surajpal Verma, Eur. J. Anal. Chem., 5, 156 (2010). Y. Zhong, Z. Jiao and Y. Yunqiu, Biomed. Chromatogr., 20, 319 (2006). J. Kuhn, C. Gotting and K. Kleesiek, Talanta, 80, 1894 (2010). J. Kuhn, C. Prante, K. Kleesiek and C. Gotting, Clin. Biochem., 42, 83 (2009). B. Shen, S. Li, Y. Zhang, X. Yuan, Y. Fan, Z. Liu, Q. Hu and C. Yu, J. Pharm. Biomed. Anal., 50, 515 (2009). M.-O. Benoit-Biancamano, P. Caron, E. Levesque, R. Delage, F. Couture and C. Guillemette, J. Chromatogr. B, 858, 159 (2007). H. Benech, S. Hascoe, V. Furlan, A. Pruvost and A. Durrbach, J. Chromatogr. B, 853, 168 (2007). M. Platzer, K. Jahn, J. Wohlrab and R.H.H. Neubert, J. Chromatogr. B, 755, 355 (2001). V. Srivatsan, A.K. Dasgupta, P. Kale, R. Verma, P. Joshi, D. Soni, M. Patel, G. Soni, J. Patel and H. Modi, J. Chromatogr. A, 1031, 259 (2004). D. Teshima, N. Kitagawa, K. Otsubo, K. Makino, Y. Itoh and R. Oishi, J. Chromatogr. B, 780, 21 (2002). U.D. Renner, C. Thiede, M. Bornhauser, G. Ehninge and H.-M. Thiede, Anal. Chem., 73, 41 (2001). K. Na-Bangchang, O. Supasyndh, T. Supaporn, V. Banmairuroi and J. Karbwang, J. Chromatogr. B, 738, 169 (2000).